8508
Instrument FSC
MOD2
This field specifies one of the following:
1. The load impedance that the AC voltage is referenced to, or
2. Resistance measurement type: HV Ohms, Normal Ohms, or True Ohms.
3. PRT accuracy data to use in the 8508A accuracy file.
• 5Z
• 6Z
• 7Z
• HI
• NO
• OC
• P1
• P2
• P3
• P4
• P5
• P6
• P7
• P8
• P9
• blank
Rules:
50 Ω
600 Ω
75 Ω
HV Ohms
Normal Ohms
True Ohms
PRT 1, User Defined (default: Fluke 8508A-PRT or Hart 5626)
PRT 2, User Defined (default: Fluke 8508A-SPRT or Hart 5699)
PRT 3, User Defined
PRT 4, User Defined
PRT 5, User Defined
PRT 6, User Defined
PRT 7, User Defined
PRT 8, User Defined
PRT 9, User Defined
(see below)
•
The MOD2 field may specify a reference impedance only when the
NOMINAL field specifies decibels.
•
The MOD2 field may specify a PRT only when the NOMINAL field specifies
temperature.
•
When MOD2 field specifies HI (HVOhms), the M8508 FSC must lock the
20 MΩ, 200 MΩ, 2 GΩ, or 20 GΩ range.
•
When MOD2 field specifies NO (Normal Ohms), the Nominal field is ≥
20 MΩ, and the M8508 FSC CON field specifies LOI (low current), the
M8508 FSC must lock the 200 MΩ or 2 GΩ range.
•
The MOD2 must be blank for any of the following conditions:
1. DC Voltage: The NOMINAL specifies voltage and the MOD1 field is
blank.
2. DC Current: The NOMINAL specifies current.
3. AC Current: The NOMINAL or MOD1 field specifies current.