BULLETIN F-68
8. MEASUREMENT
When wiring, piping settings and mounting of the sensor are completed, start the measurement.
The contents displayed on the MEASURE screen are as follows.
Instantaneous flow meter
⋅ On the MEASURE screen, instantaneous flow, instantaneous flow velocity, analog output, and analog input
are displayed.
Of the 4 stages displayed on the MEASURE screen, 2 contents can be arbitrarily allocated to the 1st and 2nd
stage. Allocation is accomplished by selection of “UNIT”.
If the flow rate is displayed when water flow stops, refer to page 9-3, “ZERO ADJUST” and page 9-5,
“CUT OFF”.
If the flow display fluctuates, refer to page 9-3 “RESPONSE SET”.
Integrated flow rate
⋅ When the integrated flow rate is displayed, “+TOTAL” and “-TOTAL” are fixed on the third stage and 4th
stage, respectively.
⋅ Integrated flow rate value is available in the range from 0000000 to 999999. If the value exceeds 9999999,
it resets to 0000000.
Battery alarm
Clock
Indicator
Instantaneous flow rate,
flow velocity, etc.
Unit of indication
value
95-06-27 11:49
How to view display
FLOW RATE
UNIT:
UNIT
m3/h
Mantissa
Exponent
0 =1 time
X10 1 =10 times
X10 2 =100 times
Example) 1.200 x101
X10
corresponds to
1.2X10=12.
MEASURE
x10
x10
VELOCITY
UNIT:
m/s
x10
+ TOTAL
UNIT:
mL
STOP
When the cursor points to
the "MEASURE"
Tab as displayed, use
care since the measurement value will not
change.
Move the cursor within
the screen by pressing
the
key.
RESET
- TOTAL
UNIT:
L/s
L/min
L/h
ML/d
m3/s
m3/min
m3/h
Mm3/d
BBL/s
BBL/min
BBL/h
MBBL/d
m/s
AI %
AO %
Instantaneous
flow rate
Instantaneous
flow velocity
Analog input
Analog output
mL
Unit of totalize
STOP
RESET
Status display of
totalize
NORMAL
Status display
Resetting of integrated flow rate
Integrated flow rate in forward direction
Integrated flow rate in reverse direction
8-1