Chromatographic Checkout 5 Phosphorus performance 6 If it is not already installed, install the phosphorus filter. 7 Create or load a method with the parameter values listed in Table 7. Table 7 FPD Checkout Conditions (P) Column and sample Type HP-5, 30 m × 0.32 mm × 0.25 µm (19091J-413) Sample FPD checkout (5188-5953) Column mode Constant pressure Column pressure 25 psi Split/splitless inlet Temperature 200 °C Split/splitless Mode Splitless Purge flow 60 mL/min Purge time 0.75 min Packed column inlet Temperature 200 °C Detector Temperature 200 °C (On) Hydrogen flow 75 mL/min (On) Air (Oxidizer) flow 100 mL/min (On) Mode Constant makeup flow OFF Makeup flow 60 mL/min (On) Makeup gas type Nitrogen Flame On Lit offset Typically 2 pA High voltage On Oven Advanced User Guide Initial temp 70 °C Initial time 0 min Rate 1 25 °C/min 73