Chromatographic Checkout
Methylparathion
Isooctane
Sulfur performance
13 Install the sulfur filter and filter spacer.
14 Make the following method parameter changes.
Table 8
Sulfur method parameters (S)
Parameter
Value ( mL/min)
H2 flow
50
Air flow
60
15 Ignite the FPD flame if not lit.
16 Display the signal output and monitor. This output
typically runs between 50 and 60 but can be as high as
70. Wait for the output to level off. This takes
approximately 1 hour.
If the baseline output is too high:
• Check column installation. If installed high, the
stationery phase burns out and increases measured
output.
• Check for leaks.
• Bake out the detector and column at 250 °C.
• Wrong flows set for installed filter.