8 • Measurement and adjustment Spark Test set-up Testbox 20100-019 Power S1 DC 70 V + - ECB connection ERBE AE NE Fig. 8-4 • The test setup is designed as shown in the illustration above. • The test specimen is connected to the power supply via the power cord. • The test specimen is switched on. • On the PC the "VIO HF Adjustment Tool" software is installed. Test procedure 1. Perform adjustment step 6 as described in the software. 2. Using the <Continue> button switch to the next test step. 3. Do not switch off the test specimen during rearrangement for next test step. 56 / 144 Art.-Nr.: 80116-271 01.07 Dok.-Nr: D010050-EN, Ver.: 007, ÄM-Nr: 09088, Gültig ab: 26.02.07, Gedruckt: CDIERL/15.09.08, Ausdruck nicht maßstäblich und kein Original. Desktop PC