Transcript
8a Software Download and Diagnostics (DU) 8a.1.2 Power Diagnostics The service application provides power supply diagnostics for the various circuit boards in the Avance anesthesia machine. Selecting Power Diagnostics on the Main Menu brings up the following menu selections in the left-hand frame and the instructions in the right-hand frame: Main Menu Power Diagnostics Gas Diagnostics Vent Diagnostics Display Diagnostics Special Functions ------------------------------------------- Power Diagnostics Power Controller Power Diagnostics Power Control Anes Control Board Mixer Select a menu item to see the power status and measured voltages. Vent Interface Bd Display Unit -> Main Menu To troubleshoot a power problem, start with the power controller and work forward. Problem voltages are in red. Software Download 1009-0357-000 09/07 (DU) 8a-3