Download Operator`s manual

Transcript
Setup Procedures: Setup Screens
Song Player
You can activate a song to be played from the monitor’s speaker to celebrate each
birth.
Song Player Volume
This field sets the volume of the song player.
SpO2 Scale
Two scale options are available for printing the %MSpO2 trend, for compatibility
with %FSpO2 trending. (%FSpO2 is always trended on a scale of 0–100%.) The
scale is printed on the paper along with the trend.
„
Auto: If only the %MSpO2 trend is printing, the trend plots on an expanded
scale of 60–100% or 50–100%, depending on the paper.* Whenever the%FSpO2
is being trended, the monitor automatically switches to a scale of 0–100% for
both %MSpO2 and %FSpO2.)
„
0–100%: This option configures the %MSpO2 trend to always plot at a fixed
scale of 0–100% — even when the maternal trend prints alone.
Paper Speed
This field selects the paper speed of the strip chart recorder.
Paper Chime
This field enables/disables an audible tone to indicate a low-paper or out-of-paper
condition. Refer to “Chapter 12, Recorder Modes”.
Recorder Light
This field turns the recorder light on and off. This light permits the room lights to be
dimmed without sacrificing visibility of the strip chart recorder. A patient can
remain resting, without the disturbance of overhead lights, while a nurse or
physician checks the strip chart.
IMPORTANT
RECORDER LIGHT—The recorder light is only available on
monitors purchased with Software Version 3.0 or greater. Older
monitors can be upgraded to Software Version 3.0 or greater;
however, the recorder light will not be available. If your monitor
has a recorder light: the LIGHT field displays on the General
Setup screen; and a frosted plastic light cover is located above the
strip chart recorder.
Paper Chime Volume
This field sets the volume of the paper chime. As you adjust the volume, a sample
tone sounds.
*
4-10
The %MSpO2 trend is plotted over a range of 60–100% on paper with a HR scale of 30–240 BPM. The
%MSpO2 trend is plotted over a range of 50–100% on paper with a HR scale of 50–210 BPM.
120 Series Maternal/Fetal Monitor
2015589-001
Revision C