Spectra Alerts: Resetting Alerts Resetting Alerts NOTE: FHR data is collected over time for analysis. Resetting an alert clears all data from the monitor’s memory for both FHR1 and FHR2. If you do not agree with an alert (see “False Pattern Recognition” and “Mode Switching”), you can clear the data being used via the Alert field on the associated setup screen (FECG or US/US2). To reset an alert: NOTE: It is possible that the Spectra Alerts feature may generate the same alert again. 1. Access the setup screen associated with the alert—indicated by the flashing FHR numerics. Select the mode title softkey (FECG or US/US2). 2. Highlight the Alert field. Whenever you display the setup screen, this field is set to ON. 3. Change the Alert field setting to RESET. (If you change the field to RESET by mistake and wish to change it back, simply set it to ON again.) 4. Once an alert is reset: audio and visual indications are removed; and the alert parameters information clears. False Pattern Recognition The system may recognize accelerations as baseline. Mode Switching During dual FHR monitoring, the system may “confuse” the FHRs following mode switches after delivery of the presenting twin. To guard against mode changes prior to delivery of the presenting twin: Use US for the presenting twin and US2 for the second twin. When switching to FECG for the presenting twin, disconnect the US connector which is no longer in use. If an alert is generated following delivery of the presenting twin, evaluate the tracing to determine if there are any clinical factors contributing to the alert. If you feel the alert was generated in error, change the Alert field on the associated setup screen to RESET. Consider the following: Revision C If you continue monitoring the second twin with US2, and you disconnect the FECG transducer which is no longer in use: the US2 fetal heart rate now moves to the primary display; and the FHR trend, which was bold, now prints using a normal print density. If you later switch to internal monitoring for the second twin and you disconnect the US2 transducer which is no longer in use: the fetal heart rate for the second twin displays as FECG in the primary heart rate area; and the FHR trend continues to print using a normal print density. 120 Series Maternal/Fetal Monitor 2015589-001 D-11