FiO2 Accuracy: ± 3% of full scale
Bypass Alarm Activation:
18 - 22 psi (1.2 - 1.5 bar)
Primary Outlet Flow Range: 3 - 30 L/min 3
Auxiliary Outlet Flow - Range: 0 - 30 L/min
Bleed Flow:at 50 psi (3.4 bar) - 3 L/min or less
Maximum Combined Flow: (all outlets)
≥ 30 L/min
No. of Primary Outlet Ports: 1/ 3 /1 DISS/ 5 DISS/ Order separately
No. of Auxiliary Outlet Ports: 1/ 1/ 1 DISS/ 1 DISS/ Order separately
Inlet Pressure:- 2.1 - 5.2 bar(30 - 75 psi)
•
Inlet Fitting: DISS /DISS/ NIST
•
Warranty Terms: 3 years warrantee & CMC contract form with charges from Year 4-6
•
•
•
•
66 T piece Resuscitator
Easy to use
The experience, training, concentration and fatigue level of the operator do not affect the pressures delivered.
Flexibility in operation
The neopuff can accept and deliver oxygen concentrations from 21 % to 100 % coming from a flowmeter or a blender. The T- piece
circuit can connect to infant resuscitation masks or endotracheal tubes.
Accurate pressure delivery
The fast acting manometer provides constant reassurance of mask seal and delivered PIP and PEEP.
Alternate temporary respiratory support
Consistent PEEP can be delivered to assist with breathing during transport or ventilator circuit change.