8.
Unscrew the two screws from the 25-pin cable connector and unplug the
connector from the manifold bracket.
Probe Air
Lines
Pneumatic Manifold
Supply Line;
One-touch
Connector
25-Pin Connector;
two 1.5mm Allen
fasteners
Access hole to the
other fastener.
Figure U-15—Location of the Pneumatic Manifold Supply Line and Wiring
Connector
U
9.
Remove the three screws securing the Pneumatic Module to the
RoadRunner chassis.
Figure U-16—Pneumatic Module fastener locations. The Probe Vacuum
Tubes have been Disconnected.
U
To identify tubing lines see
Figure U-14 and
Figure U-15.
U—20
10. Lift the Pneumatic Module enough to remove the four Probe Air Tubes
and the Socket Actuator Air Tube (one-touch fittings).
11. Lift the Pneumatic Module free from the RoadRunner chassis.