Download Service Manual -- Fluke -
Transcript
525A Service Manual 7: Low Ohms, Adjusting Low Ohms Source When calibrating the 5-400 Ω output range (High Ohms Source), the calibrator outputs two calibration points. After the output becomes stable, the values are entered at the PC. 1. Use the same test equipment listed in Table 16. 2. Use the same connections as in 6: Adjusting High Resistance Source. 3. Set the DMM to measure 4-wire ohms. 4. Type 7, then press the Enter key on the PC. The PC will display: First Calibration Point. Enter the Ohms displayed: 5. When the reading displayed on the DMM stabilizes, type the reading value on the PC and press the Enter key. After the value has been entered, the PC displays the following: Second Calibration Point. Enter the Ohms displayed: 6. When the reading displayed on the DMM stabilizes, type the reading value on the PC and press the Enter key. After the second value has been entered, the Calibrator will return to the calibration menu. 8: Hi Ohms, Adjusting High Resistance Measure When adjusting the calibration for the 5-4000 Ohms Measure range (High Ohms Measure), the Calibrator will first measure a short and then two resistances. Refer to Table 19 for a list of equipment used in this section. Table 19. List of Equipment for High Ohms Measure Calibration Adjustments Manufacturer Model Equipment Fluke FLUKE-525A/LEADS Test Lead Set Fluke 742A-10K Resistance Standard, 10K Ω Fluke 5520A Calibrator Wavetek 1281 DMM 1. Insert a 4-wire short into the 525A 4W RTD Ω jacks. 2. Type 8 then press the Enter key on the PC. The following will show on the PC: Input 0 Ohms, Press Enter When Stable 3. Allow a few seconds for stabilization, then press the Enter key. In a few moments the following appears on the PC: Input ~2000 Ohms, Enter the actual value when stable: The ~2000 indicates that the desired resistance is about 2000 Ω. The actual resistance is not as important as the accuracy with which the value is known. The method described below uses the 1281 to establish a ratio between the 5520A output and the 742 to improve the accuracy with which the 5520A’s output resistance is known. 38