Download SERVICE MANUAL
Transcript
6-5. EXECUTING IOP MEASUREMENT (5) Manual Adjust In order to execute mirror time adjustment, the following standard procedures must be followed. (1) 1. 2. 3. 4. 5. 6. 7. 8. 9. 0. In standby mode, press [TOP MENU], [CLEAR], [POWER] to enter Remocon Diagnosis Mode. Remocon Diagnosis Menu 0. 1. 2. 3. 4. 5. External Chip Check Servo Parameter Check Drive Manual Operation Emergency History Check Version information Video Level Adjustment Select “2. Drive Manual Operation” by pressing the 2 key on the remote commander. The screen will appear as below. Drive Manual Operation 1. 2. 3. 4. 5. 0. (3) Servo Control Track/Layer Jump Manual Adjustment Tray Aging Mode MIRR time adjust Return to top Menu Select “3. Manual Adjustment” by pressing the 3 key on the remote commander. The screen will appear as below. Manual Adjust 1. 2. 3. 4. 5. 6. 7. 8. 9. 0. Track Balance Adjust: Track Gain Adjust: Focus Balance Adjust: Focus Gain Adjust: Eg boost Adjust: Iop: TRV. Level: S curve(FE) Level: RFL(PI) Level: MIRR Time: ? ? Change Value [RETURN] Return to previous menu (4) Track Balance Adjust: Track Gain Adjust: Focus Balance Adjust: Focus Gain Adjust: Eq Boost Adjust: Iop. ED: TRV. Level: S curve(FE) Level: RFL(PI) Level: MIRR Time: Change Value [0] Return to previous menu Model : DPX-1790_UC2 IF-con Ver : 1.000 (5B84) Syscon Ver : 1.81 (2) Wait until a hexadecimal number appear. Select Iop by pressing 6 key on the remote commander. 6-3 (6) Convert data from hexadecimal to decimal using conversion table. (9) Press [RETURN] to return back to previous menu. (10) Press 0 to return to Top Menu and power OFF the DVD Player.