SECTION 8 SPECIFICATIONS
ENVIRONMENT
Operating: 41 to 104 F (5 to 40 C); 10 % to 90 % relative humidity
Transporting and Storage: -4 to 140 F (-20 to 60 C); 10 % to 90 % relative humidity
Atmospheric Pressure: 0 - 10,000 feet (0 - 3000 meters)
or equivalent atmospheric pressure
Relative Humidity: 10 - 90 % (104 F max)
DELIVERY RATE RANGE
Lines A and B: 0.1 to 99.9 mL/hr (in 0.1 mL/hr increments)
100 to 999 mL/hr (in 1 mL/hr increments)
Concurrent Delivery: 0.5 mL/hr minimum for each line
PlumSet: 500 mL/hr cumulative (A + B) maximum
KVO: 1 mL/hr or the last primary delivery rate, whichever is less
VTBI Range: 0.1 to 99.9 mL (in 0.1 mL/hr increments)
100 to 9999 mL (in 1 mL/hr increments)
OCCLUSION ALARM AND
LIMITS
Distal: The distal occlusion alarm sounds after the distal tubing
or set outlet fitting becomes occluded
Proximal: The proximal occlusion alarm sounds within three pumping
cycles when the tubing proximal to the cassette becomes
occluded
Distal Pressure Limit (Without
Alarm): 1 to 15 psi; maximum pressure limit is user-selectable;
factory setting is 6 psi
Maximum Infusion Pressure: 20 psi
AIR-IN-LINE ALARM
PlumSet (Distal): Bolus: 0.1 mL of air or larger
Cumulative: 0.25 mL of air out of 4.9 mL of fluid
PlumSet (Proximal): Bolus at 0.5 mL, total 1 mL (0.5 mL concurrent)
COMMUNICATION
Ethernet LAN: Shielded Ethernet cable plugged into an RJ-45 connector
Wireless LAN: Device name: Hospira MedNet Wireless 802.11 a/b/g Module
Standards: IEEE 802.11 a/b/g
Transmit Power: 802.11 b/g - 17.6 dBm; 802.11 a - 19 dBm
Antenna: Integrated Surface Mount Antenna
Certification: FCC Part 15.247, 15.407; IC RSS-210, RSS-102
FCC ID: STJ-80411396001, IC: 5627A-80411396
Ethernet IP Address: 192.168.0.100
Subnet Mask: 255.255.0.0
Plum A+ with Hospira MedNet Software
8-2
Technical Service Manual