Download HIAC ROC
Transcript
Table of contents Specifications on page 3 Data management on page 15 General information on page 4 Maintenance on page 18 Installation on page 7 Troubleshooting on page 19 User interface and navigation on page 11 Replacement parts and accessories on page 21 Operation on page 12 Specifications Specifications are subject to change without notice. Specification Details Dimensions (L x W x H) 8.9 x 10.7 x 8.9 cm (3.5 x 4.2 x 3.5 in.) Enclosure IP 66 Fitting connections SAE -4; SAE -8 Power requirements 9 to 33 VDC, 150 mA Storage temperature –40 to 85 °C (–40 to 185 °F) Operating temperature –10 to 60 °C (–14 to 140 °F) Altitude limit 2000 m (6562 ft) Overvoltage category I Pollution degree 4 Protection class III Light source Laser diode, Class I Particle size/channel 4, 6, 14 and 21 μm (ISO MTD) Storage/operating humidity 97% relative humidity, non-condensing Fluid compatibility Hydraulic and lubrication oils, mineral, synthetic (phosphate ester compatible) Fluid viscosity 2 to 424 cSt1 Reports ISO 4406, NAS 1638 and SAE AS4059 cleanliness codes Wetted materials Bronze, aluminum (anodized), steel, stainless steel, sapphire, Aflas Performance verification Optional validation certificate available (±0.5 ISO code with ISO MTD at 2.8 mg/L concentration at 100 mL/min) Weight 746 grams (2 lb) Serial interface RS232 and RS485, 9600 or 19200 Baud, 8 data bits, no parity, 1 stop bit Communication protocol MODBUS RTU Flow rate for ROC-01, ROC-11, ROC-31, ROC-41, ROC-61, ROC-71, ROC-81, ROC-91 50 to 500 mL/min (0.01 to 0.1 gal/min) Flow rate for ROC-21, ROC-51 3.8 to 38 L/min (1 to 10 gal/min) ® English 3