Download Attention
Transcript
4. Installation site/ 5. Unpacking 4. Installation site 5. Unpacking The Leica LMD6500/7000 should be used in a room that is free from dust, oil, chemical fumes and extreme humidity. Vibrations, direct sunlight and major temperature fluctuations should also be avoided at the workplace. Ambient conditions: Temperature 20°-35°C Relative humdity 20-80% up to 35°C Please compare the delivery carefully with the packing note, delivery note or invoice. We strongly recommend you keep a copy of these documents with the manual, so that you have information on the time and scope of the delivery later when ordering more equipment or when the microscope is serviced. Make sure that no small parts are left in the packing material. Some of our packing material has symbols indicating environment-friendly recycling. Microscopes in warm and humid climates need special care to prevent fungus growth. Please also read the instructions on the installation site in the microscope manual! Attention! Lamphousings and lasers must be installed at least 10 cm away from walls and other inflammable objects. ! Achtung! The Leica LMD6500/7000 must not be placed in close proximity to an air-conditioning system and it must not be exposed to any direct air draft. Note: Keep the packing material for storage and transport of the Leica LMD6500/7000 and its accessories. ! Attention! Avoid touching optical surfaces if possible. Any fingerprints on glass surfaces should be removed with a soft leather or linen cloth. Even slight traces of finger perspiration can soon attack the surfaces of optical instruments. Caution! On no account should you connect the laser and peripherals to the mains before the whole system has been assembled! 15