Emergency Message
10 Emergency Message
Emergency messages are triggered by the occurrence of a device internal malfunction
and are transmitted from the concerned application device to the other devices with
highest priority.
Emergency Message
Byte
Contents
0
1
Emergency
Error Code
Object 1003h,
Byte 0-1
2
3
4
5
6
7
Error
Register
Object
1001h
0
0
0
0
0
COB-Identifier = 080h + Node-ID
If the measuring system detects an internal error, an emergency message will be
transmitted with the error code of object 1003h (pre-defined error field) and the error
register object 1001h.
If the error disappears, the measuring system transmits an emergency message with
error code "0" (reset error / no error) and error register "0".