Download User guide - SPM Instrument
Transcript
The previous page illustrated the three basic alternatives on the ‘Settings’ menu with single line symptoms. Below are examples of multi-line symptoms. D D.S ame as B, but containing the first harmonic at BPFI plus three harmonics, altogether four possible matches (5). In this example, the match found by Leonova agrees with the obvious peaks in the spectrum: all dashed lines are on top of the largest lines (5). 5 B E 6 E.Same as C, marking the positions where BPFI and its three harmonics should be according to the calculations. In case of the first line (6), reality as reproduced by the FFT agrees with the calculation. However, the next three lines in the pattern are not quite in their calculated positions: they are beside the dashed lines (7). Please note that such near misses are normal, especially for the more rigid ‘Theoretical symptom’. Vibration itself is a continuous event, which is first digitalized during measurement and the subjected to mathematical manipulations to get the spectrum. At every step, there are tolerances, so an offset must be expected. The resolution of this spectrum is ≈ 1 Hz, so the line is offset by a single digital step. B:28 General measurement functions 7