Download CAS 750 Service Manual
Transcript
CAS 750 Monitor Service Manual MONITOR MESSAGES MONITOR MONITOR MESSAGE “Clock Battery” "Dead Battery" "Low Battery" "System Error" "Power Failure" “Set Clock” POSSIBLE CAUSE The monitor’s internal clock battery is almost discharged. The battery is fully discharged. The battery is almost discharged. An electronic failure has occurred within the monitors’ Main Board. Power was disconnected from the monitor. The monitor’s clock needs to be set. POSSIBLE SOLUTION Contact CAS Medical Systems to have the monitor serviced. Recharge the battery for at least 5 hours. At least 30 minutes of operation is available from when the message first appears. Recharge the battery as soon as possible. Contact CAS Medical Systems to have the monitor serviced. Depress the SILENCE/RESET pushbutton to clear the message. OR Turn the monitor Off, then back On to clear the message. The monitor’s time and date values are incorrect. Refer to Page 42, for information to set the Time and Date. The monitor’s internal clock battery needs to be replaced. Contact CAS Medical Systems. Table 5: Monitor Messages 21-02-0191 REV. 00 12/05 66