Download USER'S GUIDE
Transcript
Troubleshooting If you are having problems operating BladderScan® or ScanPoint®, review this list of common problems. If you do not find a solution here, contact your authorized BladderScan service center, your local BladderScan distributor or Verathon Corporation. To obtain contact information for Verathon Corporation’s Customer Service Center, see “Chapter 8. Contacting 9HUDWKRQ.” Common Problems Problem Description/ Resolution The scanning unit does not turn on This problem is usually caused by a dead or discharged battery. Charge the scanning unit for a minimum of six hours. When your computer is turned on, you can also charge the instrument using the communication cradle. If the scrolling battery icon does not appear after two hours, you will need to activate the scanning unit using the activation tool. Press the activation button with the activation tool. Once activated, replace the scanning unit in its cradle to continue charging. For more information about charging the battery and activating the scanning unit, see “Chapter 2. Introduction to BladderScan,” in this Urology Suite User’s Guide. The scanning unit is charged but will not scan If the scanning unit does not measure bladder volume when you press the Scan button, but the battery icon on the scanning unit’s LCD screen indicates that the battery has some power remaining, one of the following conditions may apply: The battery has some power remaining, but not enough to perform bladder imaging. In this case, the battery icon will show only one power segment. Simply place the scanning unit in the charging cradle or communication cradle to recharge the battery. For more information about charging the battery, see “Chapter 2. Introduction to BladderScan,” in this Urology Suite User’s Guide. You need to calibrate the scanning unit. In this case, the LCD screen displays zero as the number of days remaining until calibration and you must calibrate the scanning unit before you can continue to perform bladder volume measurements. See “Chapter 6. Inspecting and Maintaining BladderScan” for instructions on displaying the days remaining until calibration and calibrating the scanning unit. Note: This does not apply for the BVI 6300. The scanning unit beeps The scanning unit may beep in the following situations: You turn on the scanning unit. The scanning unit goes into sleep mode to conserve battery power. The battery power is low, and the battery requires recharging. In this case, the battery icon will show no power segments. Place the scanning unit in the charging cradle or communication cradle to recharge the battery. You need to calibrate the scanning unit. In this case, when you press and hold down the Top (Gender) button for five seconds, the LCD screen will display zero as the number of days remaining until calibration. For more information about calibrating the scanning unit, see “Chapter 6. Inspecting and Maintaining BladderScan,” in this Urology Suite User’s Guide. Note: This does not apply for the BVI 6300. The scanning unit performs and completes a bladder volume or calibration measurement. The scanning unit has begun or finished transmitting data to ScanPoint. The calibration procedure was successfully completed. You select or deselect the female option. You select the pre-void or post-void option. 94 Chapter 7 - Troubleshooting