Fieldbus Data-Interface
2.6.3.7
X4 Network/Fieldbus
Register [1]: Scale Status (read)
(dynamic status) Only reading is allowed.
Byte 0
Byte 1
Byte 2
Byte 3
Bit 7
DIM
Bit 6
STND
Bit 5
INZSR
Bit 4
CZERO
Bit 3
BELOW0
E6
Bit 2
Bit 1
OVL
>FSD
E1
E3
PowerFail
ActionActiv
TareActiv
CalChanged
Bit 0
ADUERR
E7
CmdError
TestActiv
Byte 0 corresponds to byte 7 in the output area, for weight errors see table.
ADUERR
>FSD
OVL
BELOW0
CZERO
INZSR
STND
DIM
Errror in analog conversion/ load cell circuitry (OR-function of bits E1,E3,E7)
gross value larger FSD (german SKE), scale range exceeded
scale overloaded, gross > FSD + Overload, Error 2
Gross weight negative (< 0-1/4 d)
Center of zero, weight within 1/4d range
Gross weight is within zeroset range
Scale is in stand still
Gross weight has exceeded the scale range (0-¼ d >weight >FSD + overload), (OR-funktion of
bits BELOW0, OVL).
Measuring signal is negative (inverse conversion), Error 7
Sense voltage missing / too low, Error 6
Measuring signal is > 36mV (no end of conversion), Error 3
Arithmetic error (overflow), Error 1
E7
E6
E3
E1
CmdEror
ActionActiv
PowerFail
TestActiv
CalChanged
TareActiv
2.6.3.8
Error during execution (cmdError), e.g. action 'taring' is not executed as there is no standstill.
In LastError (register 4) the error is specified. With ResetError register 89/89+128 or
register 121 the bit is reset again.
Action will be executed, is still processed
Power failure, will be set at each power-on. With bit ResetPWF register 85/85+128 or
register 117) 'Power failure reset' the bit PowerFail is reset.
Analog test is aktive
Calibration mode is active or calibration data have been changed.
FSD has to be read again to reset this bit.
Indicator has beeen tared
Register [4]: Calibration information, error byte (read)
Only reading is allowed.
Byte 0
Byte 1
Byte 2
Byte 3
EN-60
Bit 7
Bit 6
EXPO
UNIT
STEP
LASTERROR
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Sartorius