Transcript
Controlling Peggy page 65 Manual Clean This option is used for general manual cleaning and cleaning the manifold head. To do a manual cleaning, select Instrument and click Manual Clean. The manifold head will move to a safe position for easy access and the vacuum will turn on. See “Manifold Flush” on page 75 for the manual manifold cleaning procedure. NOTE: Please contact Protein Simple Technical Support with any questions regarding the manifold cleaning procedure. Cleanup This option is a fully automated cleaning step. The manifold head is flushed, the separation tray troughs are aspirated and washed, and any capillaries left in the trays or gripper are picked up and discarded. This option should be selected when the instrument has not been used for more than a week or if a run error occurs. Cleaning takes about eight minutes to complete. To start the protocol, select Instrument and click Cleanup. A window will appear with instructions: Peggy User Guide