Download Intelect Adanced Service Manual
Transcript
5- TROUBLESHOOTING Intelect Transport® Combo Therapy System 5.12 ULTRASOUND TESTS Equipment Required for 5.13 and 5.14 • Degassed Water. Refer to page 16 for Degassed Water Recipes. • Ohmic Instruments DT 100 UPM or DT 10 Ultrasound Power Meter. • Dissolved Oxygen Test Kit. Used to test oxygen level of degassed water. • Intelect Transport Combo Applicator. 5.13 ULTRASOUND APPLICATOR IDENTIFICATION TEST NOTE: Use any Intelect Transport Combo Applicator for this test. A. Ultrasound Applicator Identification Test Procedures 1. Without Ultrasound Applicator installed, turn unit on. 2. View the Ultrasound channel in the lower right corner of screen. It should read “Unplugged”. See Figure 5.11. 3. Connect Ultrasound Applicator into Applicator receptacle. See Figure 5.12. Watch Applicator LED while connecting to system. The LED should flash Green five times. 4. Look at the Ultrasound channel. It should read Available. See Figure 5.12. 5. Highlight Ultrasound. Press the Enter button. 6. Highlight Warming. Press the Enter button until On is displayed beside Warming. 7. Press the Back button. Turn System Off and back On with Main Power Switch. After System boots, view the Ultrasound channel, Warming should be visible. See Figure 5.13. FIGURE 5.11 FIGURE 5.12 FIGURE 5.13 23