Download Operating Instructions Type 8202 ELEMENT
Transcript
Type 8202 ELEMENT Maintenance and troubleshooting Red LED Current output OFF 4-20 mA Switched 2) OFF OFF Transistor output 4-20 mA Switched 2) 4-20 mA Switched 2) Icon Message displayed in the Info menu + "W:Impedance Ref" + "W:Imped. Glass" Possible cause Recommended action The impedance of the reference electrode is out of range. →→Go into the "Sensor" The message is displayed if the monitoring of the impedance of the reference electrode has been activated, depending on the set thresholds WARN LO and WARN HI (see chap. 9.13.2). The impedance of the pH electrode is out of range. The message is displayed if the monitoring of the impedance of the pH electrode has been activated, depending on the set thresholds WARN LO and WARN HI (see chap. 9.13.3). + "W:Temperature" The fluid temperature is out of range. The message is displayed if the monitoring of the fluid temperature has been activated, depending on the set thresholds WARN LO and WARN HI (see chap. 9.13.3). function of the Diagnostic menu to read the impedance value of the reference electrode (chap. 9.13.2). →→If necessary, clean then recalibrate the measurement sensor or replace it. →→Go into the "Sensor" function of the Diagnostic menu to read the impedance value of the pH electrode (chap. 9.13.3). →→If necessary, clean then recalibrate the measurement sensor or replace it. →→Go into the "Sensor" function of the Diagnostic menu to read the measured fluid temperature (chap. 9.13.3). →→If necessary, check whether the built-in Pt1000 is working correctly by measuring a fluid with a known temperature. →→If the Pt1000 is faulty, return the device to Bürkert. →→If the Pt1000 is not the cause of the problem, check the process. If the "PVAR" function of the "Output.TR1" and/or "Output.TR2" menus is set to "warning" (see chap. 9.11.10); else, the transistor outputs are operating depending on the set thresholds. 2) 67 English