Download BROOKFIELD CT3 TEXTURE ANALYZER Operating Instructions
Transcript
I.2 Utilities Input Voltage: Input Frequency: Power Consumption: Fuse: Power Cord Color Code: Hot (live) Neutral Ground (earth) 90-265 VAC 50/60 Hz 150 VA Two 4 amp, 5 x 20mm, Time-lag United States Outside United States Black White Green Brown Blue Green/Yellow Main supply voltage fluctuations are not to exceed ±10% of the nominal supply voltage. I.3 Units The CT3 Texture Analyzer uses the SI system of units for all parameters. Parameter Load Deformation Time Speed Work Unit Grams or Newtons Millimeter Seconds Millimeter per second MilliJoules Abbreviation g or N mm s mm/s mJ I.4 Specifications Load: Range Resolution Model Grams Grams Accuracy* CT3-100 0-100 0.01 ±0.5%FSR CT3-1000 0-1000 0.10 ±0.5%FSR CT3-1500 0-1500 0.20 ±0.5%FSR CT3-4500 0-4500 0.50 ±0.5%FSR CT3-10kg 1-10000 1.0 ±0.5%FSR FSR = Full Scale Range *This value applies to the full range of operating temperatures as defined in Section I.4 Environmental Conditions. Accuracy will be greatly improved when the instrument is operated in a stable ambient temperature. Trigger Point: Model CT3-100 CT3-1000 CT3-1500 CT3-4500 CT3-10k Range Grams 0.1-10 0.2-100 0.2-150 0.5-500 1-1000 Resolution Grams 0.01 0.10 0.2 0.5 1.0 Recommended trigger value settings are given in Table II.1 Brookfield Engineering Labs., Inc. Page Manual No. M/08-371A0708