Fig. 105: Possible transformer input voltages
6
5
4
3
2
1
0
146
– Fuse 6.3A for 220 out put voltage, Ch T
Other kinds of fuses are used in the system: the position of some of
them is shown in the following figure.
They are accessible without removing or opening the modules.
Remove the small cover on the main power plug of these system
modules to change their fuse or check that they work. Their values
are indicated in the next table.
Other fuses are present in the system modules and their values are
indicated in the following table.
Tab. 10: Fuses values and position
Module
NOTICE
The secondary voltage from the console transformer is 220V
Check if the main fuse has the correct value (depending on the
country main power supply) and check if it’s interrupted:
– Fuse 15A for 100÷110VAC supply voltage, Ch T
– Fuse 6.3A for 220÷240VAC supply voltage, Ch T
The secondary fuse already inserted in the secondary stage of the
transformer is:
Fuse value
GRA
6.3 A
SRFA
3.15 A
SINT
0.16 A
SRIB
1A
CTERM
1.6 A
10 A
4A
Ch
T
T
2A
0.8 A
T
T
4A
T
These fuses are present in the spare part installation kit.
The equipment must be permanently connected to the mains Power
supply: 100/110/220/230/240 VAC ±10%, 50/60 ±10%Hz, 1.3kW
(refer to the Site Planning Guide).
NOTICE
A TWO POLE SEPARATING SWITCH MUST
ALWAYS BE INSTALLED BETWEEN THE
MAINS AND THE SYSTEM