User Manual for Patient Monitor
4.9.5 Temperature Parameter Settings
Figure 4.28 Temperature Settings
 TEMP Hi:Temperature 1 upper alarm setting, adjustable 0.1 – 60.0°C (32.1 - 140.0°F)
Lo:Temperature 1 lower alarm setting, adjustable 0 – 59.9°C (32.0 - 139.8°F)

Unit:Temperature unit. The factory setting is °C (Celsius) and can be set to °F (Fahrenheit).
4.9.6 NIBP Parameter Settings
Figure 4.29 NIBP Settings(1)
Figure 4.30 NIBP Setting(2)
NOTE: NIBP Setup Screen will be displayed in two pages (As shown in Fig 4.29 and Fig 4.30). Move the
gray cursor to “
”, then press the navigation knob to enter the second page where three operations:
“NIBP calibration”, “Test gas leakage” and “Initial pressure” can be conducted.
 SYS Hi/Lo: Systolic pressure upper and lower alarm limits.
 DIA Hi/Lo: Diastolic pressure upper and lower alarm limits.
 MAP Hi/Lo: Mean arterial pressure upper and lower alarm limits.
 PR Hi/Lo: Pulse rate upper and lower alarm limits.
 Unit: the pressure unit. mmHg or kPa. The factory default is mmHg.