User Manual for Patient Monitor 4.9.5 Temperature Parameter Settings Figure 4.28 Temperature Settings TEMP Hi:Temperature 1 upper alarm setting, adjustable 0.1 – 60.0°C (32.1 - 140.0°F) Lo:Temperature 1 lower alarm setting, adjustable 0 – 59.9°C (32.0 - 139.8°F) Unit:Temperature unit. The factory setting is °C (Celsius) and can be set to °F (Fahrenheit). 4.9.6 NIBP Parameter Settings Figure 4.29 NIBP Settings(1) Figure 4.30 NIBP Setting(2) NOTE: NIBP Setup Screen will be displayed in two pages (As shown in Fig 4.29 and Fig 4.30). Move the gray cursor to “ ”, then press the navigation knob to enter the second page where three operations: “NIBP calibration”, “Test gas leakage” and “Initial pressure” can be conducted. SYS Hi/Lo: Systolic pressure upper and lower alarm limits. DIA Hi/Lo: Diastolic pressure upper and lower alarm limits. MAP Hi/Lo: Mean arterial pressure upper and lower alarm limits. PR Hi/Lo: Pulse rate upper and lower alarm limits. Unit: the pressure unit. mmHg or kPa. The factory default is mmHg. -33-