Download User Manual

Transcript
User Manual for Patient Monitor
9.4 Respiration Monitoring
9.4.1 Measuring Principle
The air will be filled into alveolus or be expelled during respiration, and the chest’s volume changes with this
process. Because the conductivity of air is lower than body tissues, the chest’s impedance will be changed by the
inflation. With this specialization, the respiration can be monitored through putting safe current into body and
measuring the change of voltage between the electrodes. The product will transmit the high-frequency current
whose frequency is much higher than ECG frequency (but with the safe current limit) to the ECG electrodes
(placed at the both sides of chest), which can be detect ECG signal and chest’s impedance at the same time, and
the respiratory rate will be measured through impedance method by the software. So the additional electrodes for
respiratory measurement are unnecessary.
9.4.2 Factors affecting respiration monitoring

Place the white (RA) and red (LL) electrodes on the cross, if the line between two electrodes in liver area or
heart area, the artificial error may occur.

Respiration monitoring doesn’t support monitoring the patient who do much movement, or may lead to
wrong alarm.
9.5 Temperature Monitoring
The sensor is thermo-resistor type (25
℃ 5kΩ)
the temperature of measured part through measuring the voltage. There is a period responding time, so
the accurate temperature value display after a while. The temperature monitoring can be divided into
two measuring method: measure through body surface temperature and through the temperature inside
the body cavity (placed in mouth or anus).
℃~ 37℃; inside body cavity: 36.5℃~ 3
Normal value: body surface: 36.5
Notes:

Attach the TEMP transducer to the patient; generally if the TEMP transducer and skin doesn’t contact closely,
the measured value becomes lower, so for those who have requirement for temperature, add a proper martial to
transducer and fix it with adhesive tape to make them contact firmly.

Especially for pediatric patient, they like sports, pay more attention to the transducer fixing.
9.6 CO2 Monitoring
9.6.1 Measuring Principle
The principle is based on the fact that CO2 molecules absorb infrared light energy of specific wavelengths, with the
amount of energy absorbed being directly related to the CO2 concentration. When an IR light beam is passed through a
gas sample containing CO2, the electronic signal from a photodetector (which measures the remaining light energy),
can be obtained. This signal is then compared to the energy of the IR source, and calibrated to accurately reflect CO2
concentration in the sample. To calibrated, the photodetector’s response to a known concentration of CO2 is stored in
the monitor’s memory.
The monitor determines CO2 concentration in the breathing gases by measuring the amount of light absorbed by these
gases.EtCO2 is display as a numerical value in millimeters of mercury (mmHg), percent (%), or kilopascals (kPa). In
addition, a CO2 waveform (capnogram) may be displayed which is a valuable clinical tool that can be used to assess
patient airway integrity and proper endotracheal tube placement. Respiration rate is calculated by measuring the time
interval between detected breaths.
-61-