Download User's Manual
Transcript
1. ZERO POINT ADJUSTMENT IMPORTANT 1. Do not turn off the power to the transmitter immediately after completion of the change in the LRV (and/or URV) setting(s). Note that powering off within thirty seconds after setting will ca use a return to the previous settings. 2. Changing LRV automatically changes URV to the following value. URV previous URV (new LRV previous LRV) 3. If the range-setting push-button and external zero-adjustment screw are not touched during a range-change operation, the transmitter automatically switches back to the normal operation mode. Integral indicator Range-setting switch (Push-button) Note: Use a thin bar which has a blunt tip, e.g., a hexagonal wrench, to press the rangesetting push-button. F0106.EPS Figure 1.2.1 Range-setting Switch 1-4 IM 01C22T01-01E