Download MC100 Pump Control Module User's Manual (Profibus DP)
Transcript
Users Manual MC100 8.2 Warnings handling All warnings are cleared automatically, when data within limits are received via FB, or with clear-bit: MC100 command B1.4 In warning 08-13 by the phrase individual pumps, means “real” connected pumps and pump 0 means the pseudo pump, which holds common parameters in parallel and serial mode 8.2.1 Warnings Warning no. 0 1 2 3 4 5 6 7 Warning description No pending warnings Fill volume out of range Tube number is not in table Speed setting is out of range Acceleration setting is out of range Reverse setting is out of range Density setting is out of range Calibration is out of range 8 Common calibration in parallel 9 Individual calibration in serial 10 Attempt to set common parameters to individual pump in serial mode 11 Attempt to calibrate pump 0 (fictive common pump) in individual mode MC100 Users Manual Profibus DP - rev 4 Rev. 4 Action Attempt to set parameter that is outside limits. Set new parameter. Calibration attempt for Pump 0. Calibrate pumps individually. Calibration attempt for individual pump, calibrate pump 0 only Set parameters via pump 0 (fictive common pump) or clear warning. Clear by calibrating physical connected pumps (1 to 15). Page 33 of 40