Download USER MANUAL - Astec Microflow
Transcript
5.4 CABINET MAINS POWER CONNECTION Mains power connection to the cabinet is made by the use of a standard IEC 3 pin connector assembly incorporating dual pole fusing and an isolator switch (Fig. 6). Figure 6 Mains Connector and Switch 5.5 EXHAUST FILTER SAMPLE PORT This port is provided to ease access for testing of the exhaust filter, particularly in cabinets with top boxes/ducting etc. A sample port is located on the top right hand side of the cabinet. Connection to the port is made by depressing the outermost (locking) ring and then pulling the red blanking plug from the fitting. A nylon tube of outside diameter 8 mm maximum can then be pushed in the hole to make a connection. After testing is completed remove the tube and the refit blanking plug. The zone sampled could be under positive or negative pressure, so this should be considered when connection is made. The port is useful during setting up of the cabinet after a filters change or for checking the integrity of the exhaust filter. © Bioquell 52 Royce Close, West Portway, Andover Hampshire SP10 3TS United Kingdom TM040-O&M-001 Revision 2 Page 21 of 66 Tele +44 (0) 1264 835835 Fax +44 (0) 1264 835836 Email [email protected] Web www.bioquell.com