the date the vial is opened or until the „Expiration Date“ on the bottle, whichever comes first.
• Do not use smartLAB® control solution that is past the „Expiration Date“
• The smartLAB® control solution can stain clothing. If you spill
it, wash your clothes with soap and water.
• Close the bottle tightly after use.
• Left over control solution should not be added back into the control bottle.
• Store the bottle of smartLAB® control solution at room temperature, below 30°C (86°F). Do not freeze nor refrigerate.
Running a smartLAB® control solution test
Please make sure you have all necessary items at hand when running a control solution test. You need the smartLAB®sprint meter, a
smartLAB®pro test strip, and smartLAB® control solution.
1. Switch on the meter
Take a strip out of the vial and immediately close it. Insert the strip
into the meter in the direction of the arrow on the strip. The meter
turns on automatically. „Code 888“ appears on the display and afterwards the symbol for applying blood.
In case „Code 888“, DOES NOT appear on the display, please contact your supplier (see chapter „Note on smartLAB® „NO CODE“ test strips)
21