Transcript
Chapter 3 _______________________________________________________________ Installation An overpressure in the process is necessary to create a flow through the sampling cell. Note that the pressure of the sampling cell must not differ from that of the process because dewpoint temperature changes with pressure. In dirty processes, it may be necessary to use a filter between the cooling coil and the sampling cell. One more simple way of using the sampling cell with user provided accessories is shown in the figure below.The flow through the sampling cell is controlled with the needle valve and the pressure is kept equal to that of the process. 0503-022 Figure 35 Installing the Probe in High Temperatures VAISALA _______________________________________________________________________ 51