Download Angle-Meter NT User Manual
Transcript
3 Concept The Angle-Meter NT mainly consists of four functional blocks: • The base unit with the Generator Module and one or two Detector Module(s) • The coil frame, arranged as a cube, which generates three magnetic fields • The search coil(s), placed in the 3D-centre of the generated magnetic field • The preamplifier(s) for the amplification of the search coil signal(s) The Generator Module generates three sine wave signals of different frequencies. The signals are amplified by three power amplifiers within the Generator Module. The amplified voltages are applied via the Matching-Box to the coil frame. The resulting magnetic field induces a voltage in a search coil, which is pre-processed in the preamplifier. In the Detector Module, the signal is digitized, processed by a digital signal processor and outputted as analog and digital data. 4 Base Unit The base unit of the Angle-Meter NT is designed in a modular form. It consists of the power supply, one Generator Module and one or two Detector Modules. The main switch at the base unit’s back panel disconnects the device from the AC input voltage. Figure 4-1: Base unit with the Generator Module and the Main Detector Module Figure 4-2: Base unit plus Add-On Detector Module 5 Generator Module The Generator Module generates three digitally synthesized sine wave signals of different frequencies, which are amplified by three power amplifiers with balanced outputs (field signals). These are used for the generation of the magnetic fields. The frequencies of the field signals are: f1: 80 kHz (used for the generation of the magnetic field in the X-axis) f2: 96 kHz (used for the generation of the magnetic field in the Y-axis) f3: 120 kHz (used for the generation of the magnetic field in the Z-axis) The amplitudes of the field signals are tuned at the factory to achieve the specified flux density and an optimal homogeneity of the magnetic field at the 3D-centre of the coil frame. The user may attenuate the tuned field signals in steps of 20 % down to zero. This allows to reduce the magnetic field strength, for example if extremely large search coils are used, or to turn off the magnetic field. See also the description of the Detector Module in this Manual. The Generator Module also produces an internal synchronization signal, which is used by the Detector Modules to synchronize the start of the measurements. The switch at the front panel is used to toggle the operation mode of the device between Power On and Standby, where the Standby mode means, that all internal electronics are not powered, except the small auxiliary power supply used by the front switch. Angle-Meter NT User Manual, page 8