3 Concept
The Angle-Meter NT mainly consists of four functional blocks:
•
The base unit with the Generator Module and one or two Detector Module(s)
•
The coil frame, arranged as a cube, which generates three magnetic fields
•
The search coil(s), placed in the 3D-centre of the generated magnetic field
•
The preamplifier(s) for the amplification of the search coil signal(s)
The Generator Module generates three sine wave signals of different frequencies. The signals
are amplified by three power amplifiers within the Generator Module. The amplified voltages
are applied via the Matching-Box to the coil frame. The resulting magnetic field induces a
voltage in a search coil, which is pre-processed in the preamplifier. In the Detector Module, the
signal is digitized, processed by a digital signal processor and outputted as analog and digital
data.
4 Base Unit
The base unit of the Angle-Meter NT is designed in a modular form. It consists of the power
supply, one Generator Module and one or two Detector Modules. The main switch at the base
unit’s back panel disconnects the device from the AC input voltage.
Figure 4-1: Base unit with the Generator Module and
the Main Detector Module
Figure 4-2: Base unit plus Add-On Detector Module
5 Generator Module
The Generator Module generates three digitally synthesized sine wave signals of different
frequencies, which are amplified by three power amplifiers with balanced outputs (field
signals). These are used for the generation of the magnetic fields. The frequencies of the field
signals are:
f1:
80 kHz (used for the generation of the magnetic field in the X-axis)
f2:
96 kHz (used for the generation of the magnetic field in the Y-axis)
f3:
120 kHz (used for the generation of the magnetic field in the Z-axis)
The amplitudes of the field signals are tuned at the factory to achieve the specified flux density
and an optimal homogeneity of the magnetic field at the 3D-centre of the coil frame. The user
may attenuate the tuned field signals in steps of 20 % down to zero. This allows to reduce the
magnetic field strength, for example if extremely large search coils are used, or to turn off the
magnetic field. See also the description of the Detector Module in this Manual.
The Generator Module also produces an internal synchronization signal, which is used by the
Detector Modules to synchronize the start of the measurements.
The switch at the front panel is used to toggle the operation mode of the device between
Power On and Standby, where the Standby mode means, that all internal electronics are not
powered, except the small auxiliary power supply used by the front switch.
Angle-Meter NT User Manual, page 8