Download BM5VET User Manual
Transcript
BM5VET User’s Manual In case of using blood pressure transducer Check position of transducer with a different sensitivity For other cases Perform zero-point adjustment ZERO ART: (Zero-point Adjustment) Use ZERO option to set the zero-point of Transducer. Procedures (Zero reference) 1) Close the transducer stopcock on the patient’s side. 2) Open the venting stopcock on the air side. 3) Press the knob switch on the monitor panel. 4) Draw a line with the current input data in IBP area of WAVE WINDOW according to the Wave Base Line. And accord the wave line with the data. 5) Set the data as ‘0’ on the parameter screen. 6) Check if Zero reference is carried out. (Check the pressure parameter on the message window.) 7) Close the venting stopcock on the air side. 8) Open the transducer stopcock on the patient side. The pressure value should be displayed on the pressure parameter screen in a few seconds. Troubleshootings for a case that blood pressure value is not displayed on screen Description In case of ‘out of measurement range’ Action to Take Check the measurement conditions. situation In case blood pressure transducer is Replace the damaged transducer with new damaged one Rev 2.0 9. IBP 141