R&S®SGT100A
Signal Generator Settings
External Baseband Signal - Baseband Input
"Off"
Switches Off the external I/Q input signal.
Remote command:
[:SOURce<hw>]:BBIN:STATe on page 343
I/Q-Swap
Activates swapping of the I and Q signal components, which mirrors the spectrum at
the f=0 line and inverts the sign of the frequency.
Remote command:
[:SOURce<hw>]:BBIN:IQSWap[:STATe] on page 338
Connected Device
Indicates the ID of a connected R&S Instrument or R&S device.
If supported by the transmitter, also the transmitter sample rate is shown.
Note: The connected instrument is indicated only, if the connection is properly established and signal transmission is active.
Remote command:
[:SOURce<hw>]:BBIN:CDEVice? on page 337
Sample Rate
► To access this dialog, select "SGMA-GUI > Instrument Name > BB Input > Digital
BB In > Sample Rate".
In this section you can set the sample rate of the external digital baseband signal.
Sample Rate Source
Selects the source for estimating the sample rate or defining it by the user.
"User Defined"
Enables the user to define the sample rate in the entry field value.
"Digital I/Q In"
Estimates the sample rate on the digital I/Q input and displays the
value in the value field.
Note: Estimation works best if the sample rate is close to the user
defined value.
Remote command:
[:SOURce<hw>]:BBIN:SRATe:SOURce on page 343
User Manual 1176.8674.02 ─ 05
199