CALibration:CURRent:LEVel {MINimum | MAXimum}
Before using this command, you must select the output which
is to be calibrated by using INSTrument command. This
command can only be used after calibration is unsecured. It
sets the power supply to a calibration point that is entered with
CALibration:CURRent[:DATA] command.
During calibration, two points must be entered and the low-end
point (MIN) must be selected and entered first.
CALibration:SECure:CODE <new code>
This command enters a new security code. To change the
security code, first unsecure the power supply using the old
security code. Then, enter the new code.
CALibration:SECure:STATe {OFF | ON>}, <code>
This command unsecures or secures the power supply for
calibration.
CALibration:SECure:STATe?
This command queries the secured state for calibration of the
power supply.
The returned parameter is “0” (unsecured) or “1” (secured).
CALibration:STRing <quoted string>
This command records calibration information about your
power supply.
For example, you can store such information as the calibrator,
or the calibration date etc.
CALibration:STRing?
This command queries the calibration message and returns a
quoted string.
CALibration:VOLTage[:DATA] <numeric value>
This command can only be used after calibration is unsecured.
It enters a voltage value of a selected output that you obtained
by reading an external meter. You must first select a calibration
level (CAL:VOLT:LEV) for the value being entered. Two
successive values (one for each end of the calibration range)
must be selected and entered. The power supply then
computes new voltage calibration constants. These constants
are then stored in internal memory.
105