1.5 Setting the Acquisition Conditions for Input Signal to be Measured
Inhibit
≡For operating procedure, refer to 6.4≡
Measurement can be inhibited by inputting the inhibit signal to the INHIBIT IN terminal. This is
also possible while the gate is open or during measurement following the arming delay.
Measurement can be inhibited for up to 320 seconds. In addition, it is possible to select whether
to inhibit measurement using the positive or negative edge of the inhibit signal.
Functions
Relation between the inhibit signal and gating
Mesurement is inhibited.
Inhibit signal
Gate is open.
Gate
Not measured.
Input signal
S4
S5
S1
S2
S3
Start of measurement
Measurement is resumed.
Not counted as an
event during inhibition
of measurement.
Relation between the inhibit signal and gating/external arming
Mesurement is inhibited.
Inhibit signal
Gate is open.
Gate
External arming
Not measured.
Input signal
S4
S5
S1
S2
S3
Measurement is resumed.
Start of measurement
Not counted as an
event during inhibition
of measurement.
Relation between the inhibit signal and gating/external arming/arming delay
Mesurement is inhibited.
Inhibit signal
Gate is open.
Gate
Arming delay
External arming
Not measured.
Input signal
S4
S5
S1
S2
S3
Measurement
Start of measurement
Not counted as an is resumed.
event during inhibition
of measurement.