Removing and Installing the H-ESI Probe
Installing the H-ESI Probe
H-ESI probe, rear view
H-ESI needle
high voltage connector
receptacle
Sample inlet
Grounding
union holder
Sheath liquid /
calibrant inlet
8. Connect the sheath gas fitting (blue) to the sheath gas inlet (S) on the probe manifold.
See Figure 6.
9. Connect the auxiliary gas fitting (green) to the auxiliary gas inlet (A) on the probe
manifold. See Figure 6.
10. Unplug the H-ESI vaporizer cable from the ESI interlock socket.
11. Connect the H-ESI vaporizer cable to the vaporizer cable socket on the H-ESI probe. See
Figure 6.
12. Connect the 8 kV cable to the H-ESI needle high voltage receptacle on the H-ESI probe.
See Figure 7. Tighten the locking ring on the 8 kV connector.
13. Connect the sample transfer tubing (LC line) to the grounding union. See Figure 8.