7 Maintenance, cleaning and servicing
7 Maintenance, cleaning and servicing
7.1 Maintenance
The Multichannel Calibrator model Pascal ET or Pascal ET/IS is maintenance-free.
Repairs must only be carried out by the manufacturer.
7.2
GB
Cleaning
CAUTION!
■ Before cleaning, correctly disconnect the Multichannel Calibrator model Pascal ET
or Pascal ET/IS from the pressure supply, and switch it off.
■ Clean the Multichannel Calibrator model Pascal ET or Pascal ET/IS with a moist
cloth.
■ Electrical connections must not come into contact with moisture.
For information on returning the instrument see chapter “8.2 Return”
7.3
Recalibration
DKD certificate - Certificates:
We recommend that the instrument is regularly recalibrated by the manufacturer, with time
intervals of approx. 12 months. Every factory recalibration includes, additionally, an extensive freeof-charge check of all system parameters with respect to their compliance with the specification.
The basic settings will be corrected if necessary.
WIKA Operating Instruction, Model Pascal ET and Pascal ET/IS