Model 360E Instruction Manual
Operating Instructions
FROM ANALOG I/O CONFIGURATION MENU
DIAG
PREV
The instrument attempt to automatically calibrate
the channel … then beep.
ANALOG I / O CONFIGURATION
NEXT
ENTR
DIAG AIO
EXIT
DIAG AIO CONC_OUT_2 D/A/ CAL ERROR
EXIT
AIN CALIBRATED: NO
SET> EDIT
EXIT
DIAG AIO
CONC_OUT_2 CURR-Z: 0 mV
U100 UP10 UP DOWN DN10 D100 ENTR EXIT
Press SET> to select the analog output channel
to be configured:.
DIAG AIO
DIAG AIO
U100 UP10 UP DOWN DN10 D100 ENTR EXIT
CONC_OUT_2:CURR, NO CAL
< SET SET>
EXIT
EDIT
DIAG AIO
DIAG AIO CONC_OUT_2 RANGE: CURR
<SET SET>
CONC_OUT_2 SPAN: 10000 mV
Increase or decrease the current
output by 100, 10 or 1 counts.
The resulting change in output
voltage is displayed in the upper
line.
Continue adjustments until the
correct current is measured with
the current meter.
U100 UP10 UP DOWN DN10 D100 ENTR EXIT
EDIT
EXIT
DIAG AIO
DIAG AIO CONC_OUT_2 CALIBRATED: NO
< SET
CONC_OUT_2 ZERO: 27 mV
CONC_OUT_2 ZERO: 9731 mV
U100 UP10 UP DOWN DN10 D100 ENTR EXIT
EXIT ignores the
new setting, ENTR
accepts the new
setting.
EXIT
CAL
DIAG AIO
DIAG AIO AUTO CALIBRATING CONC_OUT_2
< SET
CONC_OUT_2 CALIBRATED: YES
CAL
EXIT
If a current meter is not available, an alternative method for calibrating the current loop outputs is to connect a
250   1% resistor across the current loop output. Using a voltmeter connected across the resistor, follow the
procedure above but adjust the output to the following values:
Table 6-21: Current Loop Output Calibration with Resistor
05232 Rev B3
Full scale
Voltage for 2-20 mA
(measured across resistor)
Voltage for 4-20 mA
(measured across resistor)
0%
0.5 V
1.0 V
100%
5.0 V
5.0 V
115