Download National Instruments 32100 User's Manual
Transcript
Chapter 3 Hardware Overview shown in Figure 3-8. The trigger-level range for the direct analog channel is ±10 V in 78 mV steps for the AT-MIO-16E-1, AT-MIO-16E-2, and AT-MIO-64E-3, and ±10 V in 4.9 mV steps for the AT-MIO-16XE-10 and AT-AI-16XE-10. The range for the post-PGIA trigger selection is simply the full-scale range of the selected channel, and the resolution is that range divided by 256 for the AT-MIO-16E-1, AT-MIO-16E-2, and AT-MIO-64E-3, and divided by 4,096 for the AT-MIO-16XE-10 and AT-AI-16XE-10. Note: The PFI0/TRIG1 pin is a high-impedance input. Therefore, it is susceptible to cross-talk from adjacent pins, which can result in false triggering when the pin is left unconnected. To avoid false triggering, make sure this pin is connected to a low-impedance signal source (less than 10 kΩ source impedance) if you plan to enable this input via software. Analog Input Channels + PGIA ADC - Mux PFI0/TRIG1 Analog Trigger Circuit DAQ-STC Figure 3-8. Analog Trigger Block Diagram There are five analog triggering modes available, as shown in Figures 3-9 through 3-13. You can set lowValue and highValue independently in software. In below-low-level analog triggering mode, the trigger is generated when the signal value is less than lowValue. HighValue is unused. AT-MIO/AI E Series User Manual 3-16 National Instruments Corporation