Download Graco 313296J User's Manual
Transcript
Alarms Alarms Warm Melt alarms alert you to a problem and help prevent system shutdown or application errors. If an alarm occurs, operation may stop and the following occurs. • Light tower indication changes (if equipped) • Status bar on the display shows the alarm description Diagnose Alarms See Alarm Codes and Troubleshooting, page 36, for causes and solutions to each alarm code. Clear Alarms Alarms are cleared by the solution(s) listed in the following table or from the screen in which they appear. Refer to Alarm Codes and Troubleshooting, page 36, for details. Alarm Codes and Troubleshooting Alarm Code Alarm Problem Cause Solution Clear Alarm Fluid Control Module CB1X A - Communication Error Ram A Not Found Ram cannot communicate with Verify that power is supFCM A. plied. Alarm automatically cleared by solution. Check that CAN cables are connected. Verify that selector switch is set correctly. Replace FCM A. CB2X B - Communication Error Ram B Not Found Ram cannot communicate with Verify that power is supFCM B. plied. Alarm automatically cleared by solution. Check that CAN cables are connected. Verify that selector switch is set correctly. Replace FCM B B61X Crossover Error (Ram A) B62X Crossover Error (Ram B) Inactive ram has a Not Primed Set inactive ram to Prime Cleared from Ram Alarm alarm. mode to automatically clear screen. See Appendix A alarm. User Interface Display, There is a Runaway alarm Correct runaway condition page 93. and clear alarm on Status screen 1. There is a Drum Empty alarm. Replace empty drum with full drum to clear. 36 313296J