Download Operators Guide - iJET panel scanners
Transcript
Maintenance 8-14 11. Insert the lamp-unit all the way so the internal power connections engage with the connections on the lamp-unit. 12. Close the lamp cover – Slide the cover to the left until it CLICKS shut. The cover must always be closed before turning scanner power ON. 13. Test the replacement by turning power on to see if the lamp lights. 14. Reset the part replacement counter that triggers the warning flag Press the arrow down button to reach Reset Lamp Cnt., press down to select Yes. Confirm with Enter. Reset Lamp Cnt. > YES Enter and use the arrow Press Enter to confirm reseting the lamp part replacement counter The following illustrations show removing and inserting a lamp unit. The lamp must be positioned for removal through the iJET Panel: To replace the lamp-unit: Locate the lamp cover on the backside of the scanner. Use the iJET panel to position the lamp for removal.