Download Teledyne 3160 Computer Drive User Manual
Transcript
Appendix Appendix A.1 Specifications (Subject to change without notice.) Ranges: (Six ranges with AutoRanging on all ranges) 1: 0–1 part per million (ppm) O2 2: 0–10 ppm O2 3: 0–100 ppm O2 4: 0–1000 ppm O2 5: 0–1 % O2 6: 0–25 % O2 Accuracy: * ±2% of full scale when calibrated and operated at the same constant temperature and pressure.* ±5% of full scale over the operating temperature range once temperature equilibrium is reached.* Additional 0.04 ppm on 0-1 ppm range, generated by internal scrubber (which is necessary to maintain these specifications). Response Time: 90 % of final reading in less than 61 seconds at 25 °C Operating Temperature:0-50°C (32 -122 °F) Analog Signal Outputs: Digital Data Lines: Alarms: For percent of full scale indication: one 0–1 V dc non-isolated signal one 4–20 mA dc isolated signal For range identification: one 0–1 V dc non-isolated signal one 4–20 mA dc isolated signal One RS-232C serial interface for 2-way communications with separate host computer—for remote monitoring and control of all functions. Five user-programmable absolute-reading alarm setpoints, with Form C, SPDT relays (3A resistive). TELEDYNE ANALYTICAL INSTRUMENTS A-1