Chapter 12 Technical Specifications Pressure Control Ventilation Settings, Ranges & Resolution (Continued) Setting Range Rise Time 0.1 to 0.9 seconds %O2 21% to 100% Patient Type Adult/Pediatric Table 12-4: Pressure Control Ventilation Settings, Ranges & Resolution (Sheet 2 of 2) Non-Invasive Positive Pressure Ventilation Settings, Ranges and Resolution Resolution is one unit unless otherwise noted (cmH2O is considered numerically equivalent to hPa). Non-Invasive Positive Pressure Ventilation Settings, Ranges & Resolution Setting Range Respiratory Rate 1 to 80 Bpm EPAP 2 to 25 cmH2O (hPa) IPAP 2 to 35 cmH2O (hPa) Inspiratory Time 0.1 to 9.9 seconds (Resolution is 0.1 second) Rise Time 0.1 to 0.9 seconds Inspiratory Trigger (I-trigger) Flow Sensitivity 0.5 to 20 LPM from base flow of 3 LPM above sensitivity (Resolution is 0.1 LPM) Expiratory Trigger (E-Cycle) % Peak Flow 10 to 80% of inspiratory peak flow %O2 21% to 100% Apnea Rate 1-80 Bpm Patient Type Adult/Pediatric Table 12-5: Non-Invasive Positive Pressure Ventilation Settings, Ranges and Resolution Maximum delivered flow during pressure controlled breaths in NPPV is limited to 100 LPM for the pediatric patient type. Apnea Ventilation Apnea ventilation is considered equivalent to Assist Control with Apnea Rate in VCV and PCV. In NPPV, only mandatory breaths are delivered, triggered either at the Apnea Rate or by the patient. REF 1057983 A Respironics V200 Ventilator Operator’s Manual 12-3