Download Agilent Technologies E4440
Transcript
One-Button Measurement Functions MEASURE (Spectrum Analysis Mode) 5.1.9 Burst Power One-Button Measurement Functions The burst power measurement is an accurate method of determining the average power for the specified burst. The analyzer is set into zero-span mode, with a sweep time that captures at least one burst. For more information see “Meas Setup (Burst Power)” on page 371. The burst power measurement acquires data from the “Measured Area” above when a radio standard is chosen and when Meas Setup, Meas Method, Measured Burst Width is selected. When Meas Setup, Meas Method, Above Threshold Lvl is selected, the “Measured Area” extends the burst width delimiter lines to the start and stop points. The mean carrier power is calculated by: 1. converting each trace point amplitude from dBm into linear power 2. Adding the above amplitudes together and dividing by the number of points included in the average. 3. This value is then displayed in logarithmic form (dBm). m p ------ö ö ü ìæ æ 10 ÷ ïç ç 10 ÷ ï ÷ ïç è ø ï øï ïè n 10 log 10 í ---------------------------- ý m–n ï ï ï ï ï ï î þ å ( Pavg ) = 318 Chapter 5
Related documents
Agilent Technologies N8201A Technical data
U3700 Series User's Guide
Agilent Technologies E4406A VSA Satellite Radio User Manual
HP LaserJet 4250 User's Manual
User`s Guide
User`s/Programmer`s Reference Volume 1, Core
Agilent Technologies N9320A Technical data
CALCulate:DPSA:MARKer:PEAK:LEFT (No Query Form)
Agilent Technologies Model A.08.xx Water Dispenser User Manual
WCA230A & WCA280A Option 28 TD-SCDMA - TRS
Rockwell and Brinelll Hardness Testers
King Tester Portable Brinell Hardness Tester