COMMAND
COMMAND SYNTAX
LIST M INn PRESSURES (Model 3207 Only)
LIST L
ARGUMENTS
None
DESCRIPTION
Lists the 16 m axim um pressures stored in the MINn Variable. This is active
only when MODEL is set to 3007.
EXAMPLE
To view the m inim um pressure settings for the m odule in the exam ple of
List Ranges type:
LIST L<CR>
The DSA will return the minim um pressure settings. They could appear as
follows.
SET MIN0 -6
SET MIN1 -6
SET MIN2 -6
SET MIN3 -6
SET MIN4 -18
SET MIN5 -18
SET MIN6 -18
SET MIN7 -18
SET MIN8 -18
SET MIN9 -18
SET MIN10 -60
SET MIN11 -60
SET MIN12 -60
SET MIN13 -60
SET MIN14 -60
SET MIN15 -60
NOTE
Generally, the m axim um pressure setting is 20% greater than the full scale
of the sensor to allow a user som e over range indication.
13