Download Microtest OMNIScanner2 User guide
Transcript
OMNIScanner User Guide The following sections provide detail on the certification process. Setting the Reference A reference must be set when OMNIScanner and OMNIRemote are used as a set for the first time. After the initial reference setting, the process should be repeated periodically to ensure unit verification. OMNIScanner stores reference data for up to 5 remote units by serial number. 1. To set the reference, connect OMNIScanner and OMNIRemote using the Link Adapter and Channel Adapter* supplied with the scanner (as shown). *Use the Set Reference Adapter with OMNIScanner LT. 2. Turn OMNIScanner on. 3. Press (Setup). 4. Select Set Reference. 5. On finding OMNIRemote, the scanner reports: Acquiring Reference Data. A gauge will indicate the progress. 6. Once the reference values has been recorded, OMNIScanner displays the REMOTE INFO screen. 7. Press ESC to return to OMNIScanner’s SETUP screen. 2-3