Subsystem Commands :SYSTem:ERRor:CODE[:NEXT]? code returns the error/event code. Multiple responses are listed in the FIFO (first-in-first-out) order, separated by a comma. Response data type is NR1. If the queue is empty, the response is +0. Example :SYST:ERR:CODE:ALL? :SYSTem:ERRor:CODE[:NEXT]? Reads and removes the top item in the error/event queue, and returns the top code. Syntax :SYSTem:ERRor:CODE[:NEXT]? Query response code <newline> code returns the error/event code. Response data type is NR1. If the queue is empty, the response is +0. Example :SYST:ERR:CODE? :SYSTem:ERRor:COUNt? Returns the number of items in the error/event queue. Syntax :SYSTem:ERRor:COUNt? Query response response <newline> response returns the number of items. Response data type is NR1. If the queue is empty, the response is +0. Example :SYST:ERR:COUN? :SYSTem:ERRor[:NEXT]? Reads and removes the top item in the error/event queue, and returns the top code and message. Syntax :SYSTem:ERRor[:NEXT]? Query response response <newline> 4- 136 Agilent B2900 SCPI Command Reference, Edition 2